| Model Number |
0684-00-0294-01 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during intra-aortic balloon (iab) therapy, the iab ruptured and blood was seen in the tubing.The iab was removed the same day and a replacement was inserted the next day.There was no patient harm or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned with the membrane completely unfolded and blood on the interior and exterior of the catheter.An unknown tubing was returned over part of the catheter tubing partially stitched closed.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and a leak was detected on the membrane approximately 1cm from the rear seal measuring 0.013cm in length.The reported problem was most likely triggered by a leak which was found on the membrane.Under magnification, a whitish patch was observed around the leak.This whitish patch is the typical appearance of an abrasion mark which is caused by calcified plaque in the aorta.A non-conforming material report review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Analysis of production (3331/213) - the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis (4109/213) - the review of the historical data was performed.Trend analysis (4110/213) - the overall complaint trend data for the period jul-19 to jun-21 was reviewed.Communication/interviews: (4111/213) communication/interviews were performed to obtain all possible information.Reference complaint #: (b)(4).
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.Complaint record id #(b)(4).
|
| |
|