Model Number 0684-00-0294-01
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/08/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
 
Event or Problem Description
It was reported that during intra-aortic balloon (iab) therapy, the iab ruptured and blood was seen in the tubing.The iab was removed the same day and a replacement was inserted the next day.There was no patient harm or adverse event reported.
 
Additional Manufacturer Narrative
The product was returned with the membrane completely unfolded and blood on the interior and exterior of the catheter.An unknown tubing was returned over part of the catheter tubing partially stitched closed.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and a leak was detected on the membrane approximately 1cm from the rear seal measuring 0.013cm in length.The reported problem was most likely triggered by a leak which was found on the membrane.Under magnification, a whitish patch was observed around the leak.This whitish patch is the typical appearance of an abrasion mark which is caused by calcified plaque in the aorta.A non-conforming material report review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Analysis of production (3331/213) - the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis (4109/213) - the review of the historical data was performed.Trend analysis (4110/213) - the overall complaint trend data for the period jul-19 to jun-21 was reviewed.Communication/interviews: (4111/213) communication/interviews were performed to obtain all possible information.Reference complaint #: (b)(4).
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.Complaint record id #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEGA 7.5FR. 30CC IAB
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
MDR Report Key12122230
Report Number2248146-2021-00431
Device Sequence Number11953124
Product Code DSP
UDI-Device Identifier10607567107950
UDI-Public10607567107950
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K120868
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/11/2023
Device Model Number0684-00-0294-01
Device Catalogue Number0684-00-0292
Device Lot Number3000133098
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2021
Initial Date Received by Manufacturer 06/14/2021
Supplement Date Received by Manufacturer07/08/2021
08/25/2021
Initial Report FDA Received Date07/06/2021
Supplement Report FDA Received Date07/27/2021
08/25/2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN.
Patient Age69 YR
Patient Weight83
-
-