• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC 8CM HEAVY DUTY MEDIUM ATTACH; MOTOR, DRILL, ELECTRIC - ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC 8CM HEAVY DUTY MEDIUM ATTACH; MOTOR, DRILL, ELECTRIC - ATTACHMENT Back to Search Results
Model Number MEDIUM-HD
Medical Device Problem Code Vibration (1674)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: this device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to component failure from normal wear.Udi: (b)(4).
 
Event or Problem Description
It was reported by (b)(6) that during service and evaluation, it was determined that the attachment device had a worn bearing, exhibited vibration, missing labelling and an illegible labeling etch.It was further determined that the device failed pretest for labeling assessment and vibration.It was noted that the triangle symbol was missing.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2021.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8CM HEAVY DUTY MEDIUM ATTACH
Common Device Name
MOTOR, DRILL, ELECTRIC - ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
MDR Report Key12122977
Report Number1045834-2021-01154
Device Sequence Number17347576
Product Code HBC
UDI-Device Identifier00845384004905
UDI-Public00845384004905
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K011444
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMEDIUM-HD
Device Catalogue NumberMEDIUM-HD
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/28/2021
Initial Report FDA Received Date07/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-