• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL MM46; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL MM46; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Model Number 71335546
Medical Device Problem Code Fitting Problem (2183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a thr surgery, the r3 20 deg xlpe acet lnr 28mm x 46mm come out after trying several times in the r3 3 hole acet shell mm46.Finally, another liner and cup were used.The procedure was successfully completed with no significant delay using a smith and nephew back up device.Patient was not harmed.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).Section: h3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device could not confirm the stated failure mode.The device shows signs of damage from attempted use.Dimensional evaluation was completed on the returned device for all features related to engagement with the poly liner.During evaluation, a slight deviation was noted for the internal spherical radius profile; however, this deviation is beyond the number of decimal places required per print tolerance and gdp rounding rules.No other deviations were noted.The failure mode cannot be confirmed via dimensional evaluation.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 3 HOLE ACET SHELL MM46
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12125594
Report Number1020279-2021-05759
Device Sequence Number19715174
Product Code MBL
UDI-Device Identifier03596010598219
UDI-Public03596010598219
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K070756
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71335546
Device Catalogue Number71335546
Device Lot Number20BM14123
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/23/2021
Supplement Date Received by Manufacturer11/12/2021
Initial Report FDA Received Date07/06/2021
Supplement Report FDA Received Date11/15/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/17/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-