| Model Number |
323000 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/10/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is pending but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that 3 vet syringe 0.3ml x 29g x 12.7mm hubs separated.The following information was provided by the initial reporter : the consumer reported that when needle shields were removed the needle hubs separated. date of event: unknown.Samples: yes.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 2021-07-13.H6: investigation summary customer returned (3) loose 0.3ml u-40 bd insulin syringes.The customer reported when needle shield was removed, needle hub separated.All 3 returned syringes were examined, and all 3 exhibited a detached needle hub/shield assembly.No damage to the barrel tips was observed.A review of the device history record was completed for batch # 0174596 all inspections were performed per the applicable operations qc specifications.There was one (1) notification noted that did not pertain to the complaint.Bd was able to duplicate or confirm the customer¿s indicated failure.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Capa pr1630423 has been opened to address this issue.H3 other text : see h10.
|
| |
|
Event or Problem Description
|
|
It was reported that 3 vet syringe 0.3ml x 29g x 12.7mm hubs separated.The following information was provided by the initial reporter : the consumer reported that when needle shields were removed the needle hubs separated. date of event: unknown; samples: yes.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.A review of the device history record was completed for batch # 0174596 all inspections were performed per the applicable operations qc specifications.There was one (1) notification noted that did not pertain to the complaint.
|
| |
|
Event or Problem Description
|
|
It was reported that 3 vet syringe 0.3ml x 29g x 12.7mm hubs separated.The following information was provided by the initial reporter : the consumer reported that when needle shields were removed the needle hubs separated.Date of event: unknown.Samples: yes.
|
| |
|