| Model Number |
ETBF3616C145E |
| Medical Device Problem Codes |
Leak/Splash (1354); Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/18/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An endurant ii bifurcated stent graft was implanted in the patient for the endovascular treatment of an unknown size aaa.An mri and ct scan was taken approx.6 years and 10 months later for a separate medical issue and it was reported endograft migration was observed.The ct scan revealed that the aortic neck at renal arteries now measures 45mm.The endograft has migrated 40mm into the aortic sac and there is a type 1 endo leak with a aaa measuring 100mm it was also reported the patient was asymptomatic per the physician, the cause of the event was neck dilation due to anatomy no additional clinical sequalae were reported and the patient will be monitored.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported the device was explanted on (b)(6) 2021, the patient is reportedly doing well.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|