• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAYER MEDICAL CARE, INC. STELLANT CT SYRINGE KIT; ANGIOGRAPHIC SYRINGE KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAYER MEDICAL CARE, INC. STELLANT CT SYRINGE KIT; ANGIOGRAPHIC SYRINGE KIT Back to Search Results
Model Number 86698153
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/25/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Bayer product analysis received and examined the returned suspect syringe.Visual examination confirmed the presence of a white colored particle within the fluid path.The particle was identified as a mold gate fiber that was introduced during the manufacturing process.Our investigation determined that the particle noted in the syringe was introduced during the manufacturing process and was not detected upon receipt of the product from the supplier.A 12 month review of complaint history determined the complaint rate to be (b)(4) complaints per million (cpm).
 
Event or Problem Description
The customer reported the following: a foreign material was observed in a stellant syringe.No injury or adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STELLANT CT SYRINGE KIT
Common Device Name
ANGIOGRAPHIC SYRINGE KIT
Manufacturer (Section D)
BAYER MEDICAL CARE, INC.
1 bayer drive
indianola PA 15051
MDR Report Key12128167
Report Number2520313-2021-00023
Device Sequence Number3592078
Product Code DXT
UDI-Device Identifier00616258022086
UDI-Public(01)00616258022086
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K082905
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/20/2025
Device Model Number86698153
Device Catalogue NumberSDS-CTP-QFT
Device Lot Number8597180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/25/2021
Initial Report FDA Received Date07/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-