Investigation: visual inspection: the following observations were made during the visual inspection: no abnormalities.Permeability test: the test showed that the progav 2.0 shunt system and the pediatric prechamber are permeable.Computer control test: the computer controlled test showed the opening pressure of the progav 2.0, at a reference flow rate of 20 ml/h in a horizontal position of the valve, to be 20.00 cmh2o.This is within the specified tolerance of 20 cmh2o ± 3 cmh2o.Additionally, the shuntassistant 2.0 was tested according to standard procedure in the vertical position of the valve.At a fixed opening pressure of 20 cmh2o in the vertical position, a pressure of 20 cmh2o +4/-2 cmh2o is expected.The results indicated that at a reference flow of 20 ml/h in the vertical position, the shuntassistant 2.0 had a pressure of 20.13 cmh2o.This is within the specified tolerance of 20 cmh2o +4/-2 cmh2o.Adjustability test: the progav 2.0 was found to be adjustable to all pressure settings.The shuntassistant 2.0 is a fixed pressure valve, therefore the adjustability test is inapplicable.Braking force and brake function test: the braking force of the progav 2.0 was within the specified tolerance and the brake function operated as expected.Internal inspection of product: in order to verify whether the investigated shunt system was compromised by the known risks of hydrocephalus therapy, e.G.By a build-up of natural substances (protein, blood, or tissue particles) in the cerebrospinal fluid, we have dismantled the shunt system.After dismantling of the valves, slightly deposits were found in progav 2.0.Results: based on our investigation results, we cannot identify a malfunction of the shunt system.The valves met all specifications during the investigation.At the time of the investigation, it was not clear to us how the mentioned functional impairment occurred.No functional deviations were detected.Deposits caused by natural substances in the cerebrospinal fluid, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Small amounts of non-visible deposits/proteins might compromise the integrity of the valve.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.From our point of view, no further regulatory actions are required.
|