| Model Number |
N/A |
| Medical Device Problem Code |
Computer Software Problem (1112)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The issue was evaluated and replicated in the fmsu lab; the cause was traced to a software configuration.The issue was resolved after upgrading to version 5.7 and installing a software component that enhanced the cache management.This issue was initially identified as a non-reportable event.During a retrospective review, it was determined that this complaint should be reported in an abundance of caution.If any additional relevant information becomes available, a supplemental report will be submitted.Ref: internal complaint number comp(b)(4).
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, fujifilm medical systems u.S.A., inc.(fmsu) received a customer inquiry for assistance with synapse pacs.There was an issue noted that is described below: when the user attempts to open a report for patient a, study details for patient b are displayed.This error is easily identified since name displayed does not match.There was no patient impact, serious injury or death associated with this event.The issue is considered highly detectable by a healthcare professional; however this report is being submitted in abundance of caution.
|
| |
|
Search Alerts/Recalls
|