SMITH & NEPHEW, INC. R3 0 HOLE ACET SHELL 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
|
Back to Search Results |
|
Model Number 71331850 |
Device Problems
Connection Problem (2900); Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/16/2021 |
Event Type
Injury
|
Event Description
|
It was reported that, during a thr, a r3 0 hole acet shell 50 mm, cannot be unscrewed after implanting.It is unknown why the shell was being removed and how the procedure was finished, although it was not significantly delayed.Patient was not harmed as consequence of this problem.
|
|
Manufacturer Narrative
|
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, no relevant clinical information has been provided to performed a thorough medical investigation.Based on the information provided, it is unknown why the shell was being removed or how the procedure was completed, although there was no significant delay reported.Since there was no patient harm because of this issue, no further clinical/medical assessment is warranted at this time.Should any additional relevant clinical information be provided, this complaint will be re-assessed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include but not limited to procedural/user error, surgical complication or patient medical history.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|