• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY CARDIAC/PERITONEAL CATHETER, STANDARD, BARIUM IMPREGNATED, 90 CM; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY CARDIAC/PERITONEAL CATHETER, STANDARD, BARIUM IMPREGNATED, 90 CM; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number 43103
Medical Device Problem Codes Disconnection (1171); Fracture (1260)
Health Effect - Clinical Codes Visual Impairment (2138); Dizziness (2194); Ambulation or Postural Difficulties (2544); Cognitive Changes (2551)
Date of Event 05/19/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient received a ventricular-abdominal shunt due to hydrocephalus and recovered well after the operation.On (b)(6) 2021, the patient had sudden dizziness and unsteady walking accompanied by decreased vision.The patient went to the emergency department of the hospital.After they were admitted by the hospital for hospitalization, the symptoms were relieved by infusion of mannitol.The patient's pressure of their shunt was adjusted and reduced as well during the hospitalization which their clinical symptoms improved immediately.After discharge, the symptoms recurred in a short time, and the patient was admitted to the hospital again.After the shunt was adjusted, the patient's symptoms improved within a short time and then worsened again.Ct review was performed, and hydrocephalus worsened and consciousness reaction gradually deteriorated.Ct examination of the head, neck, chest, abdomen, and pelvis revealed that the abdominal cavity end of the shunt was dislocated.Replacement surgery of the ventricle shunt was performed on (b)(6) 2021, and during the operation, it was found that the abdominal end of the shunt was broken.After the replacement of the abdominal end of the shunt, the patient's clinical symptoms improved immediately, and they were discharged in a short time.The patient denied violence and strenuous exercise that could have caused the issue.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIAC/PERITONEAL CATHETER, STANDARD, BARIUM IMPREGNATED, 90 CM
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
5290 california ave
irvine CA 92617
MDR Report Key12129815
Report Number2021898-2021-00135
Device Sequence Number14350673
Product Code JXG
UDI-Device Identifier00643169466821
UDI-Public00643169466821
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K792005
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2021
Device Model Number43103
Device Catalogue Number43103
Device Lot NumberE12234
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2021
Supplement Date Received by Manufacturer07/12/2021
Initial Report FDA Received Date07/07/2021
Supplement Report FDA Received Date07/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age16 YR
Patient Weight40
-
-