| Model Number |
402659 |
| Medical Device Problem Code |
Premature Elective Replacement Indicator (1483)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device reached eri status approx.44 months after the implantation.Please note this icd is affected by a field safety corrective action, bio-lqc, initiated in (b)(6) 2021 and the awareness date of the event was before the fsca started.
|
| |
|
Additional Manufacturer Narrative
|
|
The icd was received for analysis and inspected.Upon receipt the device was interrogated.The interrogation could be properly performed and revealed the eri battery status.All electrical parameters were verified and showed a normal device behavior.Especially, the current consumption and the battery state of discharge were found normal and as expected.There was no indication of a device malfunction.
|
| |
|