• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG POLARCUP HEAD/LINER INSERTER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG POLARCUP HEAD/LINER INSERTER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 75017089
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/28/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during field inspection, the polarcup head/liner inserter was broken.As this was noticed upon field inspection, there was not patient involvement.
 
Additional Manufacturer Narrative
It was reported that, during field inspection, the polarcup head/liner inserter was broken.As this was noticed upon field inspection, there was not patient involvement.The device use in treatment was returned for investigation.The reported failure mode could be confirmed upon visual inspection.The device is observed to be fractured.A review of the batch record revealed no deviations from the standard manufacturing process that could have contributed to the reported issue.A review of the complaint history revealed two additional complaints reported for the batch in question.However, a batch size of 1000 pieces was produced.The additional complaint is therefore not considered to indicate a manufacturing issue related to this batch.The relationship between the reported event and the device was confirmed.There is no indication that the reported device failed to meet manufacturing specifications upon release for distribution.Normal wear and tear through repeated impaction as well as miss-use of the device are known to contribute to the reported event.Additionally, repeated steam sterilization processes could contribute to an embrittlement.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19), all devices must be inspected and controlled for proper functioning after cleaning/disinfection.Based on the conducted investigation the root cause is attributed to a component failure without any design or manufacturing related issues.The need for corrective action is not indicated.Nevertheless, smith + nephew will continue to monitor the device for similar issues.The returned device will be discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARCUP HEAD/LINER INSERTER
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
MDR Report Key12131297
Report Number9613369-2021-00316
Device Sequence Number11660822
Product Code LXH
UDI-Device Identifier07611996103151
UDI-Public07611996103151
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75017089
Device Catalogue Number75017089
Device Lot NumberF71731
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2021
Initial Date Received by Manufacturer 06/28/2021
Supplement Date Received by Manufacturer09/27/2021
Initial Report FDA Received Date07/07/2021
Supplement Report FDA Received Date09/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-