• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US LPS FEM TO SLEEVE ADAPTER +0; LPS AND S-ROM : KNEE FEMORAL ACCESSORY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US LPS FEM TO SLEEVE ADAPTER +0; LPS AND S-ROM : KNEE FEMORAL ACCESSORY Back to Search Results
Model Number 1987-06-000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Pain (1994); Limb Fracture (4518)
Date of Event 06/15/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Patient underwent a complex primary tkr on the (b)(6) 2019 following a motor vehicle accident - she received a lps distal femoral replacement with mbt revision tibia.She had a fall a week ago and sustained a periprosthetic fracture of the femur and required revision surgery.The tibial components from the primary surgery were well fixed and left in place, the femoral components were removed and replaced with new components.There was a fracture of the press fit stem at the junction of the stem/sleeve as a result of the fall.The retrieved components were sent to the bio engineering department at royal perth hospital for evaluation as per standard retrieval policy here in wa (not available for return).There were no fragments generated as a result of the breakage and all parts were retrieved.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received stated that there was no extension to the surgical time.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LPS FEM TO SLEEVE ADAPTER +0
Common Device Name
LPS AND S-ROM : KNEE FEMORAL ACCESSORY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12131399
Report Number1818910-2021-14209
Device Sequence Number11630179
Product Code JWH
UDI-Device Identifier10603295078487
UDI-Public10603295078487
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K071417
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1987-06-000
Device Catalogue Number198706000
Device Lot Number349278
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/15/2021
Supplement Date Received by Manufacturer07/22/2021
08/12/2021
Initial Report FDA Received Date07/07/2021
Supplement Report FDA Received Date07/27/2021
08/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LPS DISTAL FEM COMP XXSM RT; LPS UNIV TIB HIN INS XXSM 23MM; UNIVERSAL FEM SLV FUL POR 31MM; UNIVERSAL STEM 75X10MM FLUTED
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
-
-