| Model Number |
RA002-4545SLR |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No product has been returned for evaluation.Evaluation was performed by (b)(6).If additional relevant information is obtained, then a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a revision procedure was performed on an unknown date for an unknown reason.During a review of investigation findings from lirc evidence of pin fracture was observed on the rod, galling and corrosion.Additionally it was observed that the rod was unable to be retracted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: no product has been returned for evaluation, however, investigation of the failure revealed that corrosion, surface finish and bending stresses all contribute to pin breaking.A new pin was implemented that had 65% improved strength and significantly improved corrosion resistance.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|