• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII QUICK CONNECT HANDLE; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII QUICK CONNECT HANDLE; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71440044
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a genesis ii quick connect handle/alignment tower fails to hold the pins as expected.In addition, it was noted that a genesis ii pc p/s cam module size 5 is scratched and worn down.As this was noticed upon inspection, there was not patient involvement.
 
Additional Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The quick connect handle locking mechanism is worn, rendering the device inoperable.The device shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII QUICK CONNECT HANDLE
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12132293
Report Number1020279-2021-05805
Device Sequence Number14309688
Product Code HRY
UDI-Device Identifier03596010213860
UDI-Public03596010213860
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71440044
Device Catalogue Number71440044
Device Lot Number06CM01382
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2021
Initial Date Received by Manufacturer 06/24/2021
Supplement Date Received by Manufacturer08/03/2021
Initial Report FDA Received Date07/07/2021
Supplement Report FDA Received Date08/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-