| Model Number |
0682400550 |
| Medical Device Problem Code |
Electromagnetic Interference (1194)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
It was reported that an equipment boom electrical rotational brakes disengaged when a cauterizer device was being used during a procedure.There was patient involvement due to the issue occurring during a procedure but there was no impact to the patient and caused no surgical delay.There were no reported injuries or adverse consequences.The equipment boom was manufactured before 22apr2020 and has the older style mfr board.Although we couldn¿t confirm the root cause we suspect the issue is most likely due to the previous version of mfr capacitive board.The newer revision capacitive touch board was redesigned and released back in april 2020 to be less susceptible to emi from high frequency devices.This boom is confirmed to be within scope of field action pfa2497567 (res 87331).A stryker representative will be performing the pfa inspection and will install the current revision of mfr board and confirm that the boom is operational.If any further information is obtained around this event, a supplemental will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported the boom in or 10 is moving on its own when energy unit is being used during a procedure.There were no reported injuries or adverse consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
[supplemental 001] this boom is confirmed to be within scope of field action pfa2497567 (res 87331).A stryker representative performed the pfa inspection and installed the current revision of mfr board and confirmed that the boom is operational.
|
| |
|
Event or Problem Description
|
|
It was reported the boom in or 10 is moving on its own when energy unit is being used during a procedure.There were no reported injuries or adverse consequences.
|
| |
|