| Model Number |
U411 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter stated that an unspecified number of patient samples showed leukocyte results from the u 411 analyzer that were different from values observed upon microscopic evaluation of the samples.Results were manually edited on the u 411 analyzer to 100/ul or 500/ul.An example was provided for one patient sample which had discrepant leukocyte results.No incorrect results were reported outside of the laboratory.The sample initially resulted in an unknown value for leukocytes, but this value was likely negative.Microscopic evaluation of the sample determined it contains 21-30 lukocytes/hpf.The result on the analyzer was edited to 500/ul.The test strip lot number and expiration date were requested, but not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
As calibration and controls passed, an instrument issue is not suspected.The investigation could not identify a product problem.The cause of the event could not be determined.
|
| |
|