Model Number 72201774
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/22/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that during acl procedure, inside the patient, the biosure ha 7mm x 30mm was broken.All broken pieces were removed from the patient with tweezers.The procedure was finished with a smith and nephew backup device.Surgical delay less than or equal to 30 minutes.Patient injuries were not reported.
 
Additional Manufacturer Narrative
The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection found a broken screw with debris.An analysis of the customer provided image found part of the device in packaging for part number 72201778 lot 50823748.Based on the condition of the product material found during visual inspection it was determined that additional material testing was not required.A review of the device records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.A review of risk management files found that the reported failure was documented appropriately.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of the raw material, found that the storage protocols, material specifications, and material tests were appropriately documented.The complaint was confirmed.Factors that could have contributed to the reported event include excessive force or torque, off-axis insertion, or improper preparation of the insertion site.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOSURE HA 7MM X 30MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key12136758
Report Number1643264-2021-02188
Device Sequence Number14450293
Product Code HWC
UDI-Device Identifier03596010611703
UDI-Public03596010611703
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K080358
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/17/2024
Device Model Number72201774
Device Catalogue Number72201774
Device Lot Number50811737
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 06/27/2021
Supplement Date Received by Manufacturer08/13/2021
Initial Report FDA Received Date07/08/2021
Supplement Report FDA Received Date08/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Weight56
-
-