MAUDE Adverse Event Report: AESCULAP AG AS NUT F/FEMUR EXTENS.STEM ALL SZ.NEUTR.; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG AS NUT F/FEMUR EXTENS.STEM ALL SZ.NEUTR.; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number NR400Z
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturer evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported to aesculap inc.That an as nut f/femur extens.Stem all sz.Neutr (part # nr400z) was implanted during a primary procedure performed on (b)(6) 2017.According to the complainant, aseptic loosening of the femoral component occurred.The patient underwent a revision procedure on (b)(6) 2021.The complaint device has not been returned to the manufacturer for evaluation.No patient complications occurred as a result of the revision procedure.Concomitant devices: nr194z - as tibia offset stem d12x92mm cemented - lot # 52257685, nr865z - as enduro femur spacer distal f2 8mm - lot # 52287175, nr865z - as enduro femur spacer distal f2 8mm - lot # 51905538, nr292z - as femur extens.Stem 6° d15x77 cemented - lot # 52295198, nr884z - as enduro meniscal component f2 18mm - lot # 52318309, nb015z - as enduro femoral component cemented f2l - lot # 52357231, nb011z - as enduro tibial comp.Offset cemented t1 - lot # 52328619.
 
Event or Problem Description
No updates required.
 
Additional Manufacturer Narrative
Manufacturing site evaluation: the device was not returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation of the unit was performed and no parts were returned for failure analysis.Therefore, the investigation was not able to confirm a device issue that could be associated with the reported event.The device history records (dhr) were reviewed for all available lot numbers; the devices were found to be within specification at the time of production.All device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AS NUT F/FEMUR EXTENS.STEM ALL SZ.NEUTR.
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key12137706
Report Number2916714-2021-00128
Device Sequence Number4615552
Product Code KRO
UDI-Device Identifier04046963367732
UDI-Public4046963367732
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/21/2021
Report Date (Section F) 07/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date01/31/2026
Device Model NumberNR400Z
Device Catalogue NumberNR400Z
Device Lot Number52211649
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date06/13/2021
Device Age64 MO
Event Location Hospital
Date Report to Manufacturer06/13/2021
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/08/2021
Supplement Report FDA Received Date10/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NB011Z - AS ENDURO TIBIAL COMP.OFFSET CEMENTED T1; NB015Z - AS ENDURO FEMORAL COMPONENT CEMENTED F2L; NR194Z - AS TIBIA OFFSET STEM D12X92MM CEMENTED -; NR292Z - AS FEMUR EXTENS.STEM 6° D15X77 CEMENTED -; NR865Z - AS ENDURO FEMUR SPACER DISTAL F2 8MM - LO; NR865Z - AS ENDURO FEMUR SPACER DISTAL F2 8MM - LO; NR884Z - AS ENDURO MENISCAL COMPONENT F2 18MM - LO
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA07/08/2021
-
-