Model Number 030813
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during laparoscopic appendectomy, the device could not be fire.The surgeon then used another device to resolve the issue in order to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was returned for analysis.Visual inspection noted the instrument body displayed no abnormalities.The visual inspection of the cartridge noted the single use loading unit (sulu) was partially applied.Inspection under the microscope displayed nicks on the knife blade.Functional evaluation noted the knife blade and cam bars were reset.The approximating lever opened fully and the sulu unloaded and loaded repeatedly without difficulty.The instrument and sulu were applied to test media with acceptable results; the pushers advanced, the knife cut completely and cleanly, despite the noted knife blade damage, and the remaining staple line was properly formed.The reported condition was confirmed.The analysis noted evidence that the device was not used as intended.This issue may occur if the instrument is applied over an obstruction.The instructions included with this device provide the following guidance: when positioning the instrument on the application site ensure that no obstructions such as previously clips are incorporated into the instrument jaws.Firing over an obstruction may result in improperly formed staples.Improperly formed staples may result in leakage or disruption of the staple line.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTIFIRE ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key12137850
Report Number2647580-2021-02289
Device Sequence Number14606255
Product Code GDW
UDI-Device Identifier10884523003796
UDI-Public10884523003796
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K900129
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2025
Device Model Number030813
Device Catalogue Number030813
Device Lot NumberP0L1206
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2021
Initial Date Received by Manufacturer 06/16/2021
Supplement Date Received by Manufacturer09/29/2021
Initial Report FDA Received Date07/08/2021
Supplement Report FDA Received Date10/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-