Evaluation summary: medtronic conducted an investigation based upon all information received.The device was returned for analysis.Visual inspection noted the instrument body displayed no abnormalities.The visual inspection of the cartridge noted the single use loading unit (sulu) was partially applied.Inspection under the microscope displayed nicks on the knife blade.Functional evaluation noted the knife blade and cam bars were reset.The approximating lever opened fully and the sulu unloaded and loaded repeatedly without difficulty.The instrument and sulu were applied to test media with acceptable results; the pushers advanced, the knife cut completely and cleanly, despite the noted knife blade damage, and the remaining staple line was properly formed.The reported condition was confirmed.The analysis noted evidence that the device was not used as intended.This issue may occur if the instrument is applied over an obstruction.The instructions included with this device provide the following guidance: when positioning the instrument on the application site ensure that no obstructions such as previously clips are incorporated into the instrument jaws.Firing over an obstruction may result in improperly formed staples.Improperly formed staples may result in leakage or disruption of the staple line.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|