MAUDE Adverse Event Report: BIOMERIEUX, SA VITEK MS MYCOB/NOCARDIA KIT - 415659

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMERIEUX, SA VITEK MS MYCOB/NOCARDIA KIT - 415659 Back to Search Results
Model Number VITEK® MS MYCOBACTERIUM/NOCARDIA KIT
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Product intended use: the vitek® ms mycobacterium/ nocardia kit provides the reagents and consumables needed to process samples by protein extraction and inactivation for mycobacterium and nocardia identification using the vitek® ms system.Description of the issue: a customer from the united states reported that a bead tube from the vitek ms mycobacterium / nocardia kit (ref.415659 ¿ lot 1008403350) burst during the inactivation process.The customer reported that when performing step 3 of the inactivation process (disruption of the cells), the bead tube (tube with glass bead) used with the customer bead disruptor burst.The test was performed using biosafety level 3 practices as indicated the instruction for use.The bead disruptor instrument had an enclosure and was located inside a biological safety cabinet.Laboratory technicians wore personal protective equipment (ppe).Therefore, there was no laboratory technician exposure.
 
Additional Manufacturer Narrative
A customer in the u.S.A.Notified biomérieux that a bead tube from the vitek® ms mycobacterium / nocardia kit (ref.(b)(4) ¿ lot 1008403350) burst during the inactivation process.The bead disruptor instrument was located inside biological safety cabinet and there is a specific enclosure for it.Staff wore personal protective equipment.Customer used bead ruptor elite from omni international.Investigation: the investigator reviewed the complaints database back to november 2017, to identify any previous reports similar to this customer¿s issue.No similar complaint has been recorded linked to a breakage of beads tubes.In addition, no complaint has been recorded against the batch number (batch # 1008403350).No capas nor non-conformities were identified that could be linked with the customer¿s complaint.Next, the investigator looked at the manufacturing process of vitek ms mycob/nocardia kit batch # 1008403350: - beads are a catalog reference : qiagen powerbead tubes, glass 0.5 mm cat.No./ id: 13116-50.These are not controlled internally at reception or during the packaging process.This reference of beads is used only for the vitek ms mycob/nocardia kit (no other reference impacted).- manufacturing files verification is not performed for the impacted batch (beads are catalog reference from qiagen, introduced in our production only at the packaging step).- the internal sample for batch 1008403350 was visually controlled (100 tubes) and had no broken tubes.After investigation with r&d department, - this issue was never observed internally - no specific criteria were defined during the development related to a maximal speed value during step 3 of the deactivation protocol - equipment used by r&d were vortex genie 2 bmx p/n 280174 (max speed 2700 rpm) or biospec mini-beadbeater-1 model# c321001 (max speed 4800 oscillations/min).After investigation with our supplier qiagen, - no specification sheets for the tubes could be provided - no specific criteria were defined for this product (speed value, freezing, force withstand) but they specified that beads are ¿compatible with all bead beating instruments¿ - this kind of tubes used with power lyzer homogenizer device were described in qiagen protocols.- power lyzer homogenizer device seems very close to the customer's equipment bead ruptor elite from omni international according to this information, the bead tubes should be compatible with the device used by the customer.Product documentation: package insert 20894c - equipment specification: o vortex-type mixer with adaptor or bead beater-type homogenizer o vortex-type mixer (maximum speed).- instructions for use: o all manipulation of mycobacterium must be performed using biosafety level 3 practices before the inactivation of the sample.- step 3 mentioned by the customer: o use a vortex-type mixer with adaptor (at maximum speed) to disrupt the cells for 15 minutes, or a bead beater-type homogenizer for 5 minutes.The protocol is clearly described and biosafety recommendations are mentioned.2/ sds us version a at mycobacteria / nocardia protocol step 3, only ethanol (r1) is used.The physical and chemical hazards were clearly identified in the safety data sheet.3/ vitek ms workflow user manual 4501-2233-f (see document attached): the biosafety recommendations are stated in case of mycobacterium and nocardia preparation.The biosafety recommendations are stated in vitek ms documentation.The issue was punctual.The reference (b)(4) is the only reference impacted.No action is to be taken.Root cause: cause not established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITEK MS MYCOB/NOCARDIA KIT - 415659
Common Device Name
VITEK MS MYCOB/NOCARDIA KIT - 415659
Manufacturer (Section D)
BIOMERIEUX, SA
3 route de port michaud
la balme 38390
FR  38390
MDR Report Key12139440
Report Number9615754-2021-00182
Device Sequence Number4679123
Product Code QBN
UDI-Device Identifier03573026436971
UDI-Public03573026436971
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K181412
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/30/2021
Device Model NumberVITEK® MS MYCOBACTERIUM/NOCARDIA KIT
Device Catalogue Number415659
Device Lot Number1008403350
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/11/2021
Supplement Date Received by Manufacturer08/27/2021
Initial Report FDA Received Date07/08/2021
Supplement Report FDA Received Date09/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-