| Model Number |
AC*T-DIFF F/ASSY,FINAL TEST LEVEL, 16 PARAMETERS |
| Medical Device Problem Code |
Priming Problem (4040)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service engineer (fse) was at the customer site and confirmed the reported issue.The fse cleaned the sample syringe assembly and the probe wipe block assembly.The fse flushed the probe vacuum pathway and tightened the diluent filter connections to resolve the issue.
|
| |
|
Event or Problem Description
|
|
The customer reported that the results generated on their act diff hematology instrument for two patient samples were erroneously high.Erroneous results were reported out of the laboratory.One of the patients were initially sent to the hospital emergency room but the recommendation was reversed once the customer identified an instrument problem.No other impact to patient or treatment was reported.Additional information about the impacted patient was requested.The customer technical specialist (cts) attempted to get the details regarding the patient who went to er but was unable to get additional information from the customer.
|
| |
|