MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ENCOR ENSPIRE SYSTEM 115V; REUSABLE BIOPSY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ENCOR ENSPIRE SYSTEM 115V; REUSABLE BIOPSY Back to Search Results
Model Number E4115
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the serial number for the device was provided, a review of the device history record is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 01/2022).
 
Event or Problem Description
It was reported that the driver activated and began sampling while in the holder(sample button not being pressed) waiting for the procedure for about 20 minutes.Needle was changed, driver port was changed and it still had same issue.Device was then hooked up to an alternate system and seemed to correct itself.There was no patient contact.
 
Event or Problem Description
It was reported that the driver activated and began sampling while in the holder(sample button not being pressed) waiting for the procedure for about 20 minutes.Needle was changed, driver port was changed and it still had same issue.Device was then hooked up to an alternate system and seemed to correct itself.There was no patient contact.
 
Additional Manufacturer Narrative
H10: manufacturing review: a device history record and manufacturing review was not required as the event was determined to be expected and the investigation did not identify any manufacturing and/or service-related issues.Investigation summary: the physical device was not returned for evaluation.No photos or video were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.The root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCOR ENSPIRE SYSTEM 115V
Common Device Name
REUSABLE BIOPSY
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key12141124
Report Number2020394-2021-01294
Device Sequence Number17193415
Product Code KNW
UDI-Device Identifier00801741086113
UDI-Public(01)00801741086113
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
PMA/510(K) Number
K111100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberE4115
Device Catalogue NumberE4115
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/10/2021
Supplement Date Received by Manufacturer10/01/2021
Initial Report FDA Received Date07/09/2021
Supplement Report FDA Received Date10/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-