| Model Number |
ASHA4830-01 |
| Medical Device Problem Code |
Protective Measures Problem (3015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a knee scope surgery the ambient super multivac was connected to a quantum ii console and it showed a wand error e8.The procedure was successfully completed using a back-up device.A delay greater than 30 minutes was reported.No patient injury or other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: h3, h6: the reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection of the returned instrument shows no manufacturing abnormalities.No visual issues were observed.A functional evaluation revealed the controller generated an e7 error when the wand was connected.When used with a bypass box the device generated plasma as intended.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was an isolated event.A review of customer provided image shows a controller with e8 error on screen.The complaint was confirmed.Factors that could have contributed to the reported event include: (1) previous use.(2) disconnecting the wand from the controller after power has been turned on.No containment or corrective actions are recommended at this time.
|
| |
|