| Model Number |
DBEC-125 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Extravasation (1842); Vascular Dissection (3160)
|
| Date of Event |
06/14/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.The diamondback coronary orbital atherectomy system instructions for use warns that a vessel dissection is a potential adverse event that may occur and/or require intervention.Csi id: (b)(4).
|
| |
|
Event or Problem Description
|
|
Orbital atherectomy treatment was performed in the left main and mid left anterior descending (lad) arteries.Following treatment, a type c dissection was observed in the proximal lad.Intravascular ultrasound (ivus) revealed residual calcium.Minor extravasation was observed when the ivus catheter was removed.An echocardiogram revealed minimal effusion.Angioplasty and stent placement resolved the dissection.The patient remained in the hospital, in stable condition.
|
| |
|