• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493941832300
Medical Device Problem Codes Device-Device Incompatibility (2919); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that stent dislodgment occurred.Following deployment of a 3.00 x 16mm synergy xd drug eluting stent in the ostial circumflex artery, a 3.00 x 32mm synergy xd stent was introduced.The 3.00 x 32 stent embolized in the left anterior descending artery (lad) and became entangled in the 3.00 x 16mm stent and when the embolized stent was removed, both stents came out together.The lad was pre-dilated.Two synergy xd stents were implanted overlapping in the mid and proximal lad and another synergy xd stent was implanted in the ostial circumflex to complete the procedure.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key12142840
Report Number2134265-2021-08781
Device Sequence Number14495555
Product Code NIQ
UDI-Device Identifier08714729981008
UDI-Public08714729981008
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/02/2023
Device Model NumberH7493941832300
Device Lot Number0026820399
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/24/2021
Initial Report FDA Received Date07/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-