| Model Number |
1581-24-108 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/21/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented with minor posterior medial wear and had slight wear on his patella.Surgeon removed insert and replaced with the same size insert and decided intraop to replace the patella too with the same size.All parts were well.Doi: (b)(6) 2011.Dor: (b)(6) 2021; right knee.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|