• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC SOLO CATHETER 5F RADIOLOGY DL 135CM NITINOL GW; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC SOLO CATHETER 5F RADIOLOGY DL 135CM NITINOL GW; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/13/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of refp0114 showed one other similar product complaint(s) from this lot number.The complaints for this lot number (refp0114) have been reported from the same facility.
 
Event or Problem Description
It was reported "we had a sheath in may that the tip on it became jagged upon use." no other information provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and lot history review, applicable previous investigation(s), sample (if available), labeling, applicable manufacture records, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a damaged microintroducer sheath is confirmed; however, the exact cause is unknown.One 5.0 x 10 cm microintroducer sheath was returned for evaluation.An initial visual observation showed use residue on the returned sample.The microintroducer was observed to be slightly bent at two locations along its length, and the distal tip of the sheath was observed to be damaged and somewhat flattened.A microscopic observation revealed the distal tip of the sheath was folded and buckled in multiple locations throughout its circumference.Three points along the edges were observed to be folded in on themselves and plastic deformation was observed at the corners of these folds.While the cause of the damage observed on the introducer sheath tip is unknown, possible causes include damage during handling or use.H3 other text: evaluation findings are in section h11.
 
Event or Problem Description
It was reported "we had a sheath in may that the tip on it became jagged upon use." no other information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC SOLO CATHETER 5F RADIOLOGY DL 135CM NITINOL GW
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key12144178
Report Number3006260740-2021-02718
Device Sequence Number11815964
Product Code LJS
UDI-Device Identifier00801741028120
UDI-Public(01)00801741028120
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K072230
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number3295335
Device Lot NumberREFP0114
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2021
Initial Date Received by Manufacturer 06/17/2021
Supplement Date Received by Manufacturer08/06/2021
Initial Report FDA Received Date07/09/2021
Supplement Report FDA Received Date09/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-