MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GNS II RESURF PAT 35MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GNS II RESURF PAT 35MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 71420578
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Inflammation (1932); Pain (1994)
Date of Event 06/30/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2021 a washout of a tkr was performed due to pain and inflammation, infection was not confirmed but the patellar button was loosened and the polyethylene was also explanted.The current health status of the patient is unknown.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.The images provided were reviewed and could not confirm the stated failure mode.The clinical/medical investigation concluded that, per the complaint, no relevant clinical information will be provided for inclusion in this medical investigation.Without the requested clinical information, a thorough medical investigation cannot be rendered nor can the root cause of the reported failure be determined.Based on the information provided, a washout of a tkr was performed due to pain and inflammation.Although it was reported the infection was not confirmed, the patellar button was loose, and the polyethylene was also explanted.The impact to the patient beyond that which has already been reported cannot be determined.Should any additional medical information be provided, this complaint will be re-assessed.A complaint history review found related failures for the listed batch; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.A review of the sterilization records revealed the batch was sterilized within normal parameters.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors and/or some potential probable causes that could contribute to the reported event have been identified as abnormal motion over time, bone degeneration, fit/sizing, lack of ingrowth and/or traumatic injury.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GNS II RESURF PAT 35MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12145188
Report Number1020279-2021-05878
Device Sequence Number3976114
Product Code JWH
UDI-Device Identifier03596010208316
UDI-Public03596010208316
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2021
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71420578
Device Catalogue Number71420578
Device Lot Number18EM17522
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/30/2021
Supplement Date Received by Manufacturer08/19/2021
Initial Report FDA Received Date07/09/2021
Supplement Report FDA Received Date08/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-