| Model Number |
B1140-040 |
| Medical Device Problem Code |
Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/18/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was received.Investigation is not yet completed.A follow up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that a 14x40mm armada 35 balloon dilatation catheter (bdc) was returned to abbott vascular with the packaging torn; thus, sterility was breached.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual analysis was performed on the returned device.The reported packaging damage was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints reported from this lot.Based on the reported information, a definitive cause for the packaging damage could not be determined.It may be possible that the damage occurred at the account or during transit to the account; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|