Catalog Number 8065751761
Device Problem Excessive Heating (4030)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/09/2021
Event Type  malfunction  
Manufacturer Narrative
A sample has been received by manufacturing that has not yet been evaluated.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that during a surgery handpiece got hot.Procedure details and patient impact have not been provided.Additional information has been requested, additional information received indicating during a cataract extraction a handpiece continued to be hot.There was no system message.Handpiece was changed and surgery was completed.There was no medical intervention required.There was no patient harm.
 
Manufacturer Narrative
This complaint was opened to address a reported event of temperature high.The handpiece was confirmed to have exhibited a failure mode related to a nonconforming crystal stack.This event type is consistent with a known, previously investigated issue.It is important to note that the elevated shell temperature of the handpiece is not instantaneous as a result of this event; it gradually increases with use, and can take up to approximately 5 minutes to heat up.During product use training, the surgeon is advised to discontinue use of the handpieces if there are any indications of the handpiece reaching its end of life.Quality assurance will continue to monitor for evidence of adverse trending of reported events and take further action, as appropriate.At a minimum, this will include completing reviews of complaint event code levels on a monthly basis by the product team.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key12147450
MDR Text Key260987944
Report Number2028159-2021-00723
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657517619
UDI-Public00380657517619
Combination Product (y/n)N
PMA/PMN Number
K121555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial,Followup
Report Date 08/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/09/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065751761
Device Lot Number13MK6K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2021
Date Manufacturer Received07/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-