ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE; UNIT, PHACOFRAGMENTATION
|
Back to Search Results |
|
Catalog Number 8065751761 |
Device Problem
Excessive Heating (4030)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/09/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A sample has been received by manufacturing that has not yet been evaluated.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A customer reported that during a surgery handpiece got hot.Procedure details and patient impact have not been provided.Additional information has been requested, additional information received indicating during a cataract extraction a handpiece continued to be hot.There was no system message.Handpiece was changed and surgery was completed.There was no medical intervention required.There was no patient harm.
|
|
Manufacturer Narrative
|
This complaint was opened to address a reported event of temperature high.The handpiece was confirmed to have exhibited a failure mode related to a nonconforming crystal stack.This event type is consistent with a known, previously investigated issue.It is important to note that the elevated shell temperature of the handpiece is not instantaneous as a result of this event; it gradually increases with use, and can take up to approximately 5 minutes to heat up.During product use training, the surgeon is advised to discontinue use of the handpieces if there are any indications of the handpiece reaching its end of life.Quality assurance will continue to monitor for evidence of adverse trending of reported events and take further action, as appropriate.At a minimum, this will include completing reviews of complaint event code levels on a monthly basis by the product team.The manufacturer internal reference number is: (b)(4).
|
|