MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN® 5000 ARCTIC SUN DEVICE Back to Search Results
Model Number 50000000
Device Problems Insufficient Cooling (1130); Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/16/2021
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that nurse was getting audible alarm and marginal flow from arctic sun device. No alert message on screen. Also, patient was having a hard time getting to 33c. It had been over four hours and patient was 35c. They turned device off and back on. Water reservoir was empty alarm (04). They drained water from pads and there was 1 bar. They filled device and started therapy. Patient was around 80kg with small pads in place. Mss explained that the lack of pad coverage could be part of the reason the patient was having difficulty reaching target. Nurse would add universal pad to abdomen. Water temperature was 5c and flow rate was 2. 1lpm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN® 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key12147504
MDR Text Key260949901
Report Number1018233-2021-04140
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation
Type of Report Initial,Followup
Report Date 03/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/09/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number50000000
Device Catalogue Number50000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 07/09/2021 Patient Sequence Number: 1
-
-