Investigation summary: bd was unable to reproduce the customer¿s experience with the bactec.Retention samples were visually inspected, tested for viable contamination by sub-culturing on tsa, chocolate, sabouraud and schaedler agars plates, voltage output and gram stain.All results were satisfactory.Batch history record review did not identify any evidence for which the customer submitted the complaint.Complaint is unconfirmed based on retention samples and batch history record review.Product insert warnings and precautions sections states that prior to use, each vial should be examined for evidence of contamination such as cloudiness, bulging or depresses septum, or leakage.Vials showing evidence of contamination should not be used.Also prior to use, the user should examine the vial for evidence of damage or deterioration.Vials displaying turbidity, contamination, or discoloration (darkening) should not be used.No corrective actions were required.A cross functional team continually monitors all product complaints for trends and determines if any additional actions are necessary beyond the current investigational process.
|