• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AESPIRE VIEW; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AESPIRE VIEW; ANESTHESIA GAS MACHINE Back to Search Results
Lot Number APHV00519
Device Problem No Audible Alarm (1019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/14/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Legal manufacturer: hcs (b)(4).No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.The ge healthcare service representative performed a check of the system and confirmed the reported issue.The oxygen pressure switch was replaced to resolve the reported issue.
 
Event Description
The hospital reported that the oxygen pressure alarm did not sound during testing.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESPIRE VIEW
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
joseph seliga
3000 n grandview blvd.
waukesha, WI 
MDR Report Key12147978
MDR Text Key263382257
Report Number2112667-2021-01700
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K172045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 07/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/09/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Lot NumberAPHV00519
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/14/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/08/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-