| Model Number |
IO-300 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
Low Oxygen Saturation (2477)
|
| Date of Event |
06/10/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An internal investigation was initiated to determine the cause of the malfunction.The reason for the "system hot error" was that the vents were blocked.Blocked vents would have kept the heat from dissipating from the unit and not letting cool air into the unit to cool it.This unit is running normally and as designed.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced a unit shut down.In turn, the patient visited the emergency room.
|
| |
|
Search Alerts/Recalls
|