MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 HEMOSPLIT LONG-TERM HEMODIALYSIS CATHETER, POLYURETHANE, 14.5F

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 HEMOSPLIT LONG-TERM HEMODIALYSIS CATHETER, POLYURETHANE, 14.5F Back to Search Results
Model Number 5743730
Medical Device Problem Code Difficult or Delayed Separation (4044)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/09/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 06/2022).
 
Event or Problem Description
It was reported that prior to a catheter placement, the sheath was allegedly unable to peel.Therefore, the procedure was completed using another device.There was no patient contact.
 
Event or Problem Description
It was reported that during a catheter placement, the sheath was allegedly unable to peel.Therefore, the procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported difficulty in peeling the sheath issue as no objective evidence was provided for review.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 06/2022).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOSPLIT LONG-TERM HEMODIALYSIS CATHETER, POLYURETHANE, 14.5F
Common Device Name
DIALYSIS CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key12148514
Report Number3006260740-2021-02723
Device Sequence Number4580763
Product Code MSD
UDI-Device Identifier00801741013287
UDI-Public(01)00801741013287
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K030020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5743730
Device Catalogue Number5743730
Device Lot NumberREEX2243
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/15/2021
Supplement Date Received by Manufacturer08/11/2021
Initial Report FDA Received Date07/12/2021
Supplement Report FDA Received Date08/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-