• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TITAN SPINE, LLC ENDOSKELETON¿ TT INTERBODY SYSTEM; DRIVER, PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TITAN SPINE, LLC ENDOSKELETON¿ TT INTERBODY SYSTEM; DRIVER, PROSTHESIS Back to Search Results
Model Number 4200-1001
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding patients for tlif spinal therapy.It was reported that tlif inserter broke while inserting cage, tip of inserter stuck inside cage.No patient injury occurred.There was no patient symptoms or complications as a result of this event.
 
Additional Manufacturer Narrative
H3:part # 4200-1001; lot # n131201 visual and optical inspection confirmed the threaded portion of the inserter has broken.This type of damage is consistent with torsional overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSKELETON¿ TT INTERBODY SYSTEM
Common Device Name
DRIVER, PROSTHESIS
Manufacturer (Section D)
TITAN SPINE, LLC
6140 w. executive dr. suite a
mequon WI 53092
Manufacturer (Section G)
TITAN SPINE, LLC
6140 w. executive dr. suite a
mequon WI 53092
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key12148701
Report Number3006340236-2021-00015
Device Sequence Number4508856
Product Code HWR
UDI-Device Identifier00191375002987
UDI-Public00191375002987
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K083714
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4200-1001
Device Catalogue Number4200-1001
Device Lot NumberN131201
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2021
Supplement Date Received by Manufacturer11/03/2021
Initial Report FDA Received Date07/12/2021
Supplement Report FDA Received Date11/15/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/01/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age42 YR
Patient SexFemale
-
-