• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TITAN SPINE, LLC ENDOSKELETON TT INTERBODY SYSTEM; DRIVER, PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TITAN SPINE, LLC ENDOSKELETON TT INTERBODY SYSTEM; DRIVER, PROSTHESIS Back to Search Results
Model Number 4200-1036
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient with stenosis for spinal therapy.It was reported that the tlif inserter broke during insertion of a titan tt implant.The part of the inserter that threads into the cage broke during insertion.We had to use a hospital instrument to grab and unthread it from the cage in order to get it out. the screwdriver tip broke during insertion of an iliac screw.The bone was so hard, that it put a lot of torque and stress on the driver.I did not realize the driver has sheered off until the next day when i was putting up the set.The tip of the driver remains in the tulip head of the screw and is locked in place by the rod and set screw.  it was reported that the fragment of the driver was not removed.It is inside the screw locked in place by the rod and set screw.No additional intervention is planned or was done.
 
Additional Manufacturer Narrative
H3: part# 4200-1036; lot# y191207 after visual and optical examination, it appears that the threaded tip has been broken from the pivot pin from what appears to be overload.This is consistent with overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSKELETON TT INTERBODY SYSTEM
Common Device Name
DRIVER, PROSTHESIS
Manufacturer (Section D)
TITAN SPINE, LLC
6140 w. executive dr. suite a
mequon WI 53092
MDR Report Key12148761
Report Number3006340236-2021-00017
Device Sequence Number14579974
Product Code HWR
UDI-Device Identifier00191375030201
UDI-Public00191375030201
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K083714
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4200-1036
Device Catalogue Number4200-1036
Device Lot NumberY191207
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2021
Supplement Date Received by Manufacturer09/29/2021
Initial Report FDA Received Date07/12/2021
Supplement Report FDA Received Date09/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-