Model Number SA60WF
Medical Device Problem Codes Device Damaged by Another Device (2915); Device Handling Problem (3265)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/09/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lens was returned for evaluation.Solution is dried on the lens.The lens was cut in half with only one half of the optic with an attached haptic returned.The optic is cracked and broken from being crushed on top of a post in the lens case upon return.The cartridge was not returned for evaluation.A fold inspection could not be performed due to extensive damage to the lens.Product history records were reviewed and the documentation indicated the product met release criteria.Qualified associated products were indicated.The product investigation could not identify a root cause.The lens was returned damaged.A dimensional and fold test could not be performed due to extensive damage to the lens.The cartridge was not returned for evaluation.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A facility representative reported ''loading error, lens split defective unfolder''.The procedure was completed with an alternate intraocular lens.There was no patient harm reported.Additional information has been requested; however, further information has not been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF ASPHERIC UV ABSORBING SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key12148923
Report Number1119421-2021-01357
Device Sequence Number4816684
Product Code HQL
UDI-Device Identifier00380652332392
UDI-Public00380652332392
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2023
Device Model NumberSA60WF
Device Catalogue NumberSA60WF.065
Device Lot Number12638424
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/18/2021
Initial Report FDA Received Date07/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DUOVISC VISCOELASTIC SYSTEM; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D
Patient Age72 YR
-
-