| Model Number |
SA60WF |
| Medical Device Problem Codes |
Device Damaged by Another Device (2915); Device Handling Problem (3265)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lens was returned for evaluation.Solution is dried on the lens.The lens was cut in half with only one half of the optic with an attached haptic returned.The optic is cracked and broken from being crushed on top of a post in the lens case upon return.The cartridge was not returned for evaluation.A fold inspection could not be performed due to extensive damage to the lens.Product history records were reviewed and the documentation indicated the product met release criteria.Qualified associated products were indicated.The product investigation could not identify a root cause.The lens was returned damaged.A dimensional and fold test could not be performed due to extensive damage to the lens.The cartridge was not returned for evaluation.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported ''loading error, lens split defective unfolder''.The procedure was completed with an alternate intraocular lens.There was no patient harm reported.Additional information has been requested; however, further information has not been received.
|
| |
|