Model Number AP2
Medical Device Problem Code Installation-Related Problem (2965)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/11/2021
Type of Reportable Event Death
Additional Manufacturer Narrative
According to the analysis provided by the distributor, the missed transmission of the flags associated to the test results was caused by the wrong configuration of the stago sta r max interface module by the distributor service personnel who installed the wrong driver, the driver required to interface dms with stago sta r instead of the driver required for stago sta r max.The configuration error could have been detected by the service personnel during the tests after the installation or by the users since the analyzer name is displayed on the graphic user interface.The documentation made available to the service personnel for the interface module configuration and driver installation has been reviewed and confirmed to be clear and complete.The d-dimer test result is usually interpreted along with other test results to obtain the overall clinical picture.The d-dimer test result by itself should not lead to any therapeutic decision.A test result of 0 is not believable and should trigger a rerun in the validation phase.According to the evaluation of the haematologist on site the patients death was not caused by the automation system erroneous results transmission.According to the current information, the automation system has been confirmed not to be the cause of the incident.The investigation has been done based on the information provided by the distributor.The distributor has not provided yet inpeco with the log files necessary to confirm their analysis.
 
Event or Problem Description
The automation system data management software (dms) has sent to the upper middleware incorrect d-dimer test results associated to 13 sample tubes.The analyzer sta r max, manufactured by stago, sent to dms the d-dimer test results > 20 with the flag m (max), dms did not transmit the flag and reported to the upper middleware the out of range results as 0.According to the information received from the field 3 of these patients passed away.However the laboratory haematologist specified that the d-dimer results did not cause the patients' death.
 
Additional Manufacturer Narrative
The initial report (3010825766-2021-00008) has been submitted on july 12, 2021.Supplemental report: per fda request, the product code has been corrected to dap.("fibrinogen/fibrin degradation products assay") which is the test product code for which the inaccurate results were generated.Additional information received from the distributor about the deceased patients.The initial reporter submitted the mdr 3007494875-2021-00001 on 21-jul-2021.No additional findings regarding the investigation have been communicated by the distributor.
 
Additional Manufacturer Narrative
The initial report (3010825766-2021-00008) has been submitted on july 12, 2021.A supplemental report (3010825766-2021-00008_s1) has been submitted on august 2, 2021.Supplemental report: no additional findings regarding the investigation have been communicated by the distributor.The investigation has been completed with the confirmation that the automation system design is adequate.No further actions are foreseen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APTIO AUTOMATION
Common Device Name
LABORATORY AUTOMATION SYSTEM
Manufacturer (Section D)
INPECO SA
via torraccia 26
novazzano, 6883
SZ  6883
MDR Report Key12150638
Report Number3010825766-2021-00008
Device Sequence Number20019257
Product Code DAP
UDI-Device Identifier07640172342008
UDI-Public(01)07640172342008(11)200408
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K121012
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberAP2
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/02/2021
Supplement Date Received by Manufacturer07/20/2021
08/30/2021
Initial Report FDA Received Date07/12/2021
Supplement Report FDA Received Date08/02/2021
09/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
-
-