U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WALLFLEX DUODENAL; STENT,METALLIC,EXPANDABLE,DUODENAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WALLFLEX DUODENAL; STENT,METALLIC,EXPANDABLE,DUODENAL Back to Search Results
Model Number M00565020
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Discomfort (2330)
Date of Event 06/15/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The complainant was unable to provide the suspect device lot number.Therefore, the manufacture and expiration dates are unknown.(b)(4).The complainant indicated that the device remains implanted; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation on june 21, 2021 that a wallflex duodenal stent was implanted to treat a 4cm malignant gastric outlet obstruction in the duodenum/pylorus during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2021.The patient's anatomy was tortuous.On (b)(6) 2021, the patient experienced discomfort and an endoscopy procedure was performed where it was noted that the stent migrated.The physician attempted to re-position the stent using rat tooth forceps; however, the physician was unsuccessful.A different sized wallflex duodenal stent was placed stent-in-stent to complete the procedure.The patient was prescribed npo (nothing by mouth) for two days to address the discomfort.The patient's condition was reported to be fully recovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX DUODENAL
Common Device Name
STENT,METALLIC,EXPANDABLE,DUODENAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key12150954
Report Number3005099803-2021-03409
Device Sequence Number11950015
Product Code MUM
UDI-Device Identifier08714729456490
UDI-Public08714729456490
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K062750
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00565020
Device Catalogue Number6502
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/21/2021
Initial Report FDA Received Date07/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-