| Model Number |
1521-90-052 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2006, the patient underwent the surgery with the stem, sleeve, head.The cup is other company¿s product.On (b)(6) 2021, the revision surgery will be performed due to dislocation.The liner, head and stem will be removed.No further information is available.On (b)(6) 2021, the revision surgery was completed successfully.The sleeve was left.The head and stem was replaced with the new head and stem (stem: 900573210, head: 152190051).The surgeon also tried to replace the cup side, but did not replace the cup side because of the risk of invasion when removing the cup.The surgeon commented that it would have been more effective if the cup could have been removed.The surgeon had difficulty in removing the stem because the stem was long type.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|