Model Number 2700
Medical Device Problem Codes Gradient Increase (1270); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/17/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional manufacturer narrative: bioprosthetic tissue valves can deteriorate with time and eventually fail contributing to regurgitation and/or stenosis.Structural valve deterioration (svd) is the most common reason for bioprosthesis explants and encompasses multiple failure modes, including calcification, non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes may occur singularly or concomitantly.Alternatively, nonstructural dysfunction (nsvd) may also play a role in the development of valvular regurgitation and/or stenosis.Multiple requests for additional details regarding this event (i.E., medical records) have been performed; however, no additional information has been provided.The subject device is also unavailable for evaluation as the it remains implanted in the patient.Therefore, the reported event could not be confirmed.There is currently insufficient information to determine the root cause of this event.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.If any additional information is received, a supplemental report will be submitted accordingly.
 
Event or Problem Description
It was reported that a patient with a 25mm 2700 pericardial valve, implanted in the aortic position, underwent a valve-in-valve procedure after an implant duration of 16 years, three (3) months due to unknown reason.The tavr was performed with a 26mm 9750tfx transcatheter valve.
 
Additional Manufacturer Narrative
H10: additional manufacturer narrative: stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient related contributing factors.Structural valve deterioration (svd) is the most common reason for bioprosthesis explants and encompasses multiple failure modes, including calcification, non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes may occur singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, the nonstructural dysfunction (nsvd) may also play a role in the development of valvular stenosis.The device history record (dhr) review was completed, and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.H11: corrected data: based on the additional information obtained, this event is no longer considered reportable, and this correction is being submitted.Corrected h6 device code(s) by replacing code-3190 with code-1270.
 
Event or Problem Description
It was reported that a patient with a 25mm 2700 pericardial valve, implanted in the aortic position, underwent a valve-in-valve procedure after an implant duration of 16 years, three (3) months due to stenosis.The tavr was performed with a 26mm 9750tfx transcatheter valve.It was reported that the patient is doing well.As reported, the surgical valve, performed as intended.The physician did not indicate device failure.No additional information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT AORTIC PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
MDR Report Key12155108
Report Number2015691-2021-04063
Device Sequence Number14380839
Product Code DYE
UDI-Device Identifier00690103021615
UDI-Public00690103021615
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2005
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2008
Device Model Number2700
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/17/2021
Supplement Date Received by Manufacturer07/21/2021
Initial Report FDA Received Date07/12/2021
Supplement Report FDA Received Date07/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age64 YR
-
-