Model Number F2/8TA23
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/17/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Report inconclusive.The device has been returned and is still pending their evaluation results.The user manual contains the following warning ¿do not use excessive pressure, such as bending or prying, on attachments or dissecting tools.This may cause tool to bend or break and cause injury to patient, operator and/or operating room staff.¿ we will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during a craniotomy and removal of flap procedure, the tool was broken into three pieces.It was reported that the surgeon was able to remove the two broken pieces and the third one was suctioned.It was also reported that there was no x-ray procedure performed as all the broken pieces were removed, there was no procedure delay, and the procedure was completed with the backup product.It was reported that there was no patient impact.On follow-up, it was confirmed that the patient had no further issues post-surgery.
 
Additional Manufacturer Narrative
Report confirmed.Evaluation determined that the tool was received used and broken closer to the distal end of the tool.It was also noted that the fracture site is not consistent with failure from fatigue or high load which is typically located at tool shoulder.There were no markings near the location of distal bearing which indicates excessive side loading.Additionally, the tool was broken into 3 pieces at a single fracture site which is mostly caused by high strain rate.The most likely cause of failure was using the tool in a prying motion.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGEND TAPERED DISSECTING TOOL
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
MDR Report Key12155217
Report Number1625507-2021-00243
Device Sequence Number19810879
Product Code HBE
UDI-Device Identifier00643169766259
UDI-Public00643169766259
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K020069
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/14/2026
Device Model NumberF2/8TA23
Device Catalogue NumberF2/8TA23
Device Lot NumberH5676783
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2021
Supplement Date Received by Manufacturer07/16/2021
Initial Report FDA Received Date07/12/2021
Supplement Report FDA Received Date08/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-