Model Number 72203270
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/25/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference case(b)(4).
 
Event or Problem Description
It was reported that, during the hip distraction process, the metal inner sheath of the fine traction broke off into the threads.It was unable to provide traction using the fine traction black handle.There was a backup device available.A delay of more than 30 min was reported.No patient injury or further complications were reported.
 
Additional Manufacturer Narrative
The reported device, used in treatment was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Additional Manufacturer Narrative
The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection found a discolored and torn ball joint cover and a missing black cap on the fine positioning cross bar.The threads for the fine positioning crank mechanism had a spring-like metal wire wrapped around the threads between the crank and the boot attachment piece.A functional evaluation found the fine positioning mechanism would not retract away from the bed attachment point direction.The boot attachment freely moved along the threaded shaft until it hit the spring-like piece of metal in the threads.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The complaint was confirmed.Factors that could have contributed include excess use of force inconsistent with normal use.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AHTB UNIVERSAL HIP DISTRACTOR
Common Device Name
APPARATUS, TRACTION, NON-POWERED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
MDR Report Key12155901
Report Number3003604053-2021-00231
Device Sequence Number4260075
Product Code HST
UDI-Device Identifier00885554024821
UDI-Public00885554024821
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72203270
Device Catalogue Number72203270
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/25/2021
Supplement Date Received by Manufacturer08/17/2021
10/04/2021
Initial Report FDA Received Date07/12/2021
Supplement Report FDA Received Date08/31/2021
10/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-