| Model Number |
PS1251CS |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was not provided, a review of the device history records could not be performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure, the cartridge allegedly jammed and the seeds were not able to be released from the cartridge.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure, the cartridge allegedly jammed and the seeds were not able to be released from the cartridge.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the device history record review was performed for the affected lot number of the mick cartridge.No issues were noted.Investigation summary: the mick cartridge with 6 seeds was returned together with the mick applicator.The mick cartridge was jammed in the applicator and the seeds were not able to be discharged.With the help of pliers, moderate force , pulling and rotating the cartridge was separated from the applicator.There were 6 seeds inside the mick cartridge among one bent seed.The investigation is confirmed for reported issue, and the root cause is undetermined.Labeling review: labeling was reviewed and found to be adequate.There is a caution statements, which state: "do not overtighten the round magazine head", "do not let seeds drop into the magazine groove" and "do not use force on seeds; handle seeds gently" h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|
| |
|