|
|
| Model Number |
71366052 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during a tka, a r3 mis x-bar broke.No pieces fell inside the patient.The procedure was finished without delay using a smith and nephew back up device.It is unknown if any pieces feel into the patient.No injuries were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d4, h4.H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device is missing a bar, rendering the device inoperative.The missing bar was not returned with device.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.B5.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a tka, a r3 mis x-bar broke.No pieces fell inside the patient.The procedure was finished without delay using a smith and nephew back up device.No injuries were reported.
|
| |
|
Search Alerts/Recalls
|
|
|