| Catalog Number |
2H8401 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device manufacturer address: (b)(4).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported a clearlink system non-dehp solution set line was "splitting horizontally" which resulted in a leak of an unknown chemotherapy drug.It was further reported that once the line started to split, the blue slide would push the tubing and cause the alarm to go off and when the pump was opened, saline or chemotherapy drug would spill out.This issue was found while attached to the patient.The device was used with flogard pump.The chemotherapy drug was reported as "infliximab, nivolumab or pembrolizumab.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added to h4 and h6.H10: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|