| Model Number |
8888145059P |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); Unintended Radiation Exposure (4565)
|
| Date of Event |
06/16/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the patient had a catheter placement on june 11.On june 16, the patient stated that when she was at her residence, the catheter caught on blankets and she pulled it out.The nursing home called the hospital to tell them and she came on the next day, june 17, to have it replaced.Before the surgery, when the doctor palpated the chest of the patient, he could feel a portion of the catheter that was still present.The doctor was thinking that the catheter broke.The catheter was not repaired and had no leak.There was no luer adapter issue.No other products being utilized with the device.There was no blood leakage from the broken catheter.There was no blood loss or symptoms reported and no blood transfusion was required.There were no complications.It was said that an x-ray was done to check the remaining portion of the catheter.Then, the doctor performed a cutdown (to make it easier to remove) and surgically removed the portion of the catheter (most of it actually) and replaced it with a new one with zero issues or additional treatments needed.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the patient had a catheter placement on (b)(6).On (b)(6), the patient stated that when she was at her residence, the catheter caught on blankets and she pulled it out.The nursing home called the hospital to tell them and she came inthe next day, (b)(6), to have it replaced.Before the surgery, when the doctor palpated the chest of the patient, he could feel a portion of the catheter that was still present.The doctor was thinking that the catheter broke.The catheter was not repaired and had no leak.There was no luer adapter issue.No other products being utilized with the device.There was no blood leakage from the broken catheter.There was no blood loss or symptoms reported and no blood transfusion was required.There were no complications.It was said that an x-ray was done to check the remaining portion of the catheter.Then, the doctor performed a cutdown (to make it easier to remove) and surgically removed the portion of the catheter (most of it actually) and replaced it with a new one with zero issues or additional treatments needed.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted the cannula had broken into two pieces.It was reported that there was a break on the catheter shaft.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|