| Model Number |
PL579T |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported to aesculap ag that a challenger ti-p ml-ligat.Clips 12 cartr (part # pl579t) was used during a procedure performed on an unknown date.According to the complainant, the clip pin broke and detached into the patient.The part was not able to be recovered.The fragment remained in situ.Additional information has been requested but not yet received as of this report.Additional patient information is not available.The malfunction is filed under aag reference (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 4(5) x probability of occurrence 1(5)) according to din en iso 14971 is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based upon the investigations results a capa is not necessary.
|
| |
|
Search Alerts/Recalls
|