Model Number PL579T
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported to aesculap ag that a challenger ti-p ml-ligat.Clips 12 cartr (part # pl579t) was used during a procedure performed on an unknown date.According to the complainant, the clip pin broke and detached into the patient.The part was not able to be recovered.The fragment remained in situ.Additional information has been requested but not yet received as of this report.Additional patient information is not available.The malfunction is filed under aag reference (b)(4).
 
Additional Manufacturer Narrative
Investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 4(5) x probability of occurrence 1(5)) according to din en iso 14971 is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based upon the investigations results a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHALLENGER TI-P ML-LIGAT.CLIPS 12 CARTR.
Common Device Name
LIGATION / VESSEL CLIPS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key12159154
Report Number9610612-2021-00503
Device Sequence Number14381010
Product Code FZP
UDI-Device Identifier04046963356040
UDI-Public4046963356040
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K081031
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPL579T
Device Catalogue NumberPL579T
Device Lot Number52657392
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/22/2021
Supplement Date Received by Manufacturer09/15/2021
Initial Report FDA Received Date07/13/2021
Supplement Report FDA Received Date10/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-